![]() ![]() |
Contents | EBJ Home | Single molecule measurements and biological motors |
An important issue in making single molecule measurements is precise calibration of position detectors (such as quadrant photodiodes) and active components such as piezoelectric stages and acousto-optic deflectors. There are many different ways of calibrating a setup, but a typical calibration chain might work as follows:
Additionally, the power spectrum can be used to provide an independent calibration method, i.e. one that is not dependent on the stage or AOD calibrations.
![]() |
AOD Triangle wave used to calibrate or cross-check detector. The upper panel shows the input waveform, a 1.25 μm, 3 Hz triangle wave imposed on the trap position. The lower panel shows the detector response with a 1 μm bead in the trap. Note that the low frequency ensures that the effect of viscous drag on the bead position is minimised, so that there is negligible hysteresis in the detector response. Ideally also the measurement is made at high trap stiffness to minimise the thermal motion of the bead. The amplitude used deliberately exceeds the linear range of the detector (see next figure). |
![]() |
Linearity of Quadrant Photodiode detector. This plot shows the analysis of the records in the previous figure. The detector response over 1000 nm of trap motion is shown, with successive cycles averaged together to reduce the effect of thermal motion and other noise sources. The detector shows good linearity over approximately the central 600 nm of the range; a linear fit to this portion of the data confirms that the detector and AOD calibrations cross-check correctly. |
![]() |
Contents | Next ![]() |